Description
- Input for RTD, TC and mA
- Active / passive mA output via the same two terminals
- 1 or 2 channels
- EMPHASIS assessed instrument for nuclear industry
- SIL 2-certified via Full Assessment
- Configuration and monitoring by way of detachable display front (PR 4500); process calibration and signal simulation.
- Copying of the configuration from one device to others of the same type via the display front.
- TC inputs can use either the internal CJC or a terminal with a built-in Pt100 sensor (PR 5910, channel 1 / PR 5913, channel 2) for higher accuracy.
- Advanced monitoring of internal communication and stored data.
- SIL 2 functionality is optional and must be activated in a menu point.
- 1 green and 2 red front LEDs indicate operation status and malfunction.
- 2.6 kVAC galvanic isolation between input, output and supply.
- Can be supplied separately or installed on power rail, PR type 9400.
Configure
Specifications
Specifications
Brand
Approvals & Certifications
- ATEX KEMA 07ATEX0148 X
- EAC Ex TR-CU 012/2011
- EAC LVD TR-CU 004/2011
- EAC TR-CU 020/2011
- RoHS 2011/65/EU
- ATEX 2014/34/EU
- LVD 2014/35/EU
- EMC 2014/30/EU
- SIL 2 Certified & Fully Assessed acc. to IEC 61508
- ClassNK TA18527M
- DNV Marine TAA00000JD
- EAC Ex RU C-DK.HA65.B.00355/19
- cULus, UL 61010-1 E314307
- cFMus FM19US0059X / FM19CA0032X
- IECEx KEM 09.0052X
Environmental Protection
- IP20
Humidity
- < 95 % Relative Humidity, Non-Condensing
Operating Temperature
- -20° to 60° C
Storage Temperature
- -20° to 85° C
Calibration Temperature
- 20° to 28° C
Mounting
- DIN Rail
Vibration
- ± 1 mm at 2 to 13.2 Hz
- ± 0.7 g at 13.2 to 100 Hz
Supply Voltage
- 19.2-31.2 Vdc
Response Time
- 0.4 to 60 s
Accuracy
- Better than 0.1 % of Sel. Range
RTD Type
- Pt10/20/50/100/200/250/300/P t400/500/1000, Ni50/100/120/1000
Output
- Signal Range: 0-23 mA
- Programmable Signal Ranges: 0-20/4-20/20-0/20-4 mA
Input Signal
- 0-20 mA, 4-20 mA
Applications
Applications
- 9113A-EMP can be mounted in the safe area or in zone 2 / Class I, Division 2, Groups A, B, C, D.
- Conversion and scaling of temperature (Pt, Ni and TC) and active current signals.
- 9113A-EMP has been designed, developed and certified for use in SIL 2 applications according to the requirements of IEC 61508.
- Suitable for the use in systems up to Performance Level “d” according to ISO-13849.
Documentation
Documentation
Product Manuals
Certifications
- ATEX certificate pdf 2.3 MB
- ClassNK certificate pdf 214 KB
- DNV Marine Certificate pdf 215 KB
- EAC Declaration pdf 355 KB
- EAC Ex Certificate pdf 4.6 MB
- EAC LVD Declaration pdf 890 KB
- EU Declaration of Conformity pdf 117 KB
- FM Certificate pdf 277 KB
- IECEx Certificate pdf 156 KB
- ISO 13849 Declaration pdf 150 KB
- Metrology Pattern Approval (MPA) pdf 8.0 MB
- SIL Certificate pdf 127 KB
- UL Certificate pdf 393 KB