Description
- RTD, TC, Ohm, or mV input
- Extremely high measurement accuracy
- 1.5 kVAC galvanic isolation
- Programmable sensor error value
- For DIN form B sensor head mounting
- Within a few seconds the user can program PR5331A to measure temperatures within all ranges defined by the norms.
- The RTD and resistance inputs have cable compensation for 2-, 3- and 4-wire connection.
- Continuous check of vital stored data for safety reasons.
Configure
Specifications
Specifications
Brand
Accuracy
- ± 0.05 %
Approvals & Certifications
- ATEX 2014/34/EU KEMA 10-ATEX-0002 X
- CCOE P337392 / 1
- DNV Marine Standard F. Certificate No. 2.4
- EAC TR CU 020/2011
- EMC 2014/30/EU
- IECEx dEK 13.0035X
- INMETRO DEKRA 13.0001 X
Calibration
- 20° to 28° C (68° to 82° F)
Current
- 0.2 mA
Environmental Protection
- IP00 (Encl./Terminal)
- IP68 (Encl./Terminal)
Humidity
- < 95 % Relative Humidity, Non-Condensing
Load
- ≤ (Vsupply-7.2) / 0.023 ohms
Output
- 4-20 mA Signal Range
Response Time
- 1 to 60 seconds Programmable
RTD Type
- Linch R
- Ni100
- Pt100
Supply Voltage
- 7.2-35 Vdc
Temperature Range
- -40° to 85° C (-40° to 185° F)
Vibration
- ± 1.6 mm
- ± 4 G
Applications
Applications
- Linearized temperature measurement with Pt100...Pt1000, Ni100...Ni1000, or TC sensor.
- Conversion of linear resistance variation to a standard analog current signal, for instance from valves or Ohmic level sensors.
- Amplification of a bipolar mV signal to a standard 4...20 mA current signal.